We fabricate a microelectromechanical systems (MEMS)-based device configuring the tunable air gap Fabry–Pérot filter (FPF) with a static gradient thickness filter on the same platform. The proposed double filter configuration offers a wavelength calibration approach that accurately estimates the air gap dimension in the tunable air gap FPF. The wavelength calibration is performed by utilizing the spectrally-selective and spatially-resolved transmission characteristics of the tunable air gap FPF and the static gradient thickness filter, respectively. The MEMS-compatible chip-level integration of the static gradient thickness filter facilitates device miniaturization to enable its use in handheld devices.
No takes yet. Share an insight, caveat, or question.
Sahani et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: