Device evaluation demonstrates accurate continuous blood pressure monitoring in hypertensive individuals, highlighting the potential for cuffless cardiovascular tracking.
The escalating global health impact of hypertension necessitates cutting-edge approaches for accurate and continuous blood pressure monitoring. Traditional cuff-based methods and pulse-transit-time (PTT) measurements face challenges such as discontinuity and frequent calibration. Here, we introduce a paradigm-shifting method using a flexible ultrasonic array for calibration-free estimation. The conformal ultrasonic device with piezoelectric composites, copper electrodes, and a stretchable Polydimethylsiloxane substrate, precisely measures radial artery diameter with errors controlled within 3%. The calibration-free algorithm, rooted in the Powalowshi and Pensko model, establishes a dynamic correlation, estimation errors falling within 1.2 ± 1.0, between arterial blood pressure and vessel diameter. Long-term monitoring over 30-day showcased the model's stability. Our approach eliminates the need for frequent calibration, enhances accuracy, and provides a comprehensive understanding of the dynamic relationship between blood pressure and vessel diameter. This innovation holds promise for revolutionizing continuous blood pressure monitoring, contributing to early detection and intervention in hypertensive individuals.