To observe fine blood vessels as uneven tomographic images at a high contrast, we performed tentative experiments on embossed x-ray computed tomography (CT). We constructed a cone-beam CT scanner and carried out dual-energy CT with tube voltages of 60 and 100 kV. X-ray photons penetrating through an object were detected using an indirect-conversion flat panel detector, and radiograms were produced and sent to a personal computer to reconstruct tomograms. Embossed CT was performed using pixel-shifted dual-energy subtraction, and blood vessels filled with Gd medium were observed as uneven images at high contrast and spatial resolutions. Using 1.4-time magnification imaging in combination with a 0.1-mm-focus x-ray tube, the diameter of the object visible on the embossed CT was below 100 μm.