As Cherenkov emissions are related to dose, they can be used as signal for dosimetric purposes using polarized imaging. However, when using this method, angular corrections must be applied because of Cherenkov anisotropy. Several Cherenkov light source-to-camera distances were tested to characterize its impact on the deviations from reference values of the calculated dose distributions. Average relative differences ranging from −1.6% to −11% between Cherenkov-based results and scintillation results were extracted. It is observed that an increase in Cherenkov light source-to-camera distance decreases the difference between the Cherenkov optical signal and the dose. This reduces the importance of angular corrections at higher distances (4 m and beyond).