The microfabrication of cesium vapor cells for optically pumped magnetometry relies on optimisation of buffer gas pressure in order to maximise atomic coherence time and sensitivity to external magnetic signals. We demonstrate post-bond nitrogen buffer gas pressure tuning through localized heating of an integrated micro-pill dispenser. We characterize the variation in the intrinsic longitudinal relaxation rate, γ₁₀, and magnetic sensitivity, as a function of the resulting nitrogen buffer gas pressure. Measurements are conducted through employing an optically pumped magnetometer operating in a free-induction-decay configuration. γ₁₀ is extracted across a range of nitrogen pressures between ~60-700Torr, measuring a minimum of 140Hz at 115Torr. Additionally, we achieve sensitivities as low as 130fT/√Hz at a bias field amplitude of ~ 50{μ}$T. With the optimal nitrogen buffer gas pressure now quantified and achievable post-fabrication, these mass-producible cells can be tailored to suit a variety of sensing applications, ensuring peak magnetometer performance.
No takes yet. Share an insight, caveat, or question.
McWilliam et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: