This study describes a new method for accurately measuring baloxavir marboxil (BXM) in human plasma using LC-API-MS/MS. We employed baloxavir marboxil d4 as an internal standard for precise quantification. BXM and the internal standard were isolated from plasma using tert-butyl methyl ether (TBME) extraction. The LC system separated BXM and the internal standard on a Zorbax SB C18column, and tandem mass spectrometry (MS/MS) detected them with minimal interference from human plasma components. The method showed excellent linearity (R²≥0.9956) across a wide concentration range (10.610–1229.080ng/mL). BXM stability was evaluated under various conditions, including room temperature, injector storage, freeze-thaw cycles, and long-term storage at -70°C. The method was validated according to USFDA guidelines, demonstrating its effectiveness for quantifying BXM in human plasma for pharmacokinetic studies.
No takes yet. Share an insight, caveat, or question.
Bonthu et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: