Passive samplers have long been used to measure atmospheric pollutants in both indoor and outdoor environments. They are simple to operate, and can now monitor several chemical species. However, their use is limited because they usually require a long exposition time and provide a mean value that cannot control or evidence expected or non-expected events of environmental significance. A new apparatus specifically developed for exposing Analyst© passive samplers has been used to monitor ozone and nitrogen dioxide by automatically selecting a sampling duration of 8 h, as most legislation requires. The instrument was designed to accumulate ozone or NO2 in one passive sampler for 8 h over each day, and in another passive sampler for the remaining hours. This allows for a long-time accumulation of the 8 h ozone or nitrogen dioxide in a dedicated sampler. Measurements were carried out NE of Rome at a rural site. A description of the experiments is given, with special emphasis on the quality controls. Very low uncertainties and good comparability of the data with the reference methods were obtained for both pollutants.