Answers are grounded in this paper and cite their sources.
Objective To evaluate whether automatically adjusting the volume of automated external defibrillator (AED) audio guidance to match the environmental noise level made hearing in noisy environments easier for AED users. Methods A pilot usability study was conducted. We included three emergency life-saving technicians and nine university students who assigned medical support to spectators at the Japan Rugby League One match. Participants first used the AED without a volume-change function (no-change) and then used the AED with a volume-change function (change) in a noisy environment. The primary outcome was subjective audibility (7-point Likert scale). Secondary outcome was recognition of audio guidance (yes/no). Results The mean point in subjective audibility was 4.83 in change and 2.50 in no-change; high subjective audibility was shown in change (difference, 2.33; 95% confidence interval, 1.02 to 3.64). The number of participants who were able to recognize audio guidance was 11 in the change group and six in the no-change condition, indicating that a large number of participants were able to recognize audio guidance (κ=0.36; 95% confidence interval, –0.16 to 0.88). Conclusion The results of this study indicate that this function increased subjective audibility, and 91.2% of participants recognized the AED's audio guidance. In the future, it will be necessary to develop and disseminate AEDs that are easy to use in any environment.
No takes yet. Share an insight, caveat, or question.
Nakagawa et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: