The copper electrode modified by bacteria immobilised by a polycaprolactone film was successfully developed by electropolymerisation for the purpose of determining the presence of phenol. Electrochemical techniques such as square-wave voltammetry (SWV), cyclic voltammetry (CV) and electrochemical impedance spectroscopy (EIS) were used to characterize the electrochemical properties of the Cu-polymer/bacteria electrode. The results show that the intensity of the phenol oxidation peak increases with concentration, allowing us to obtain good analytical results with DL of 2.156 × 10 –7 M and QL which is 7.2 × 10 –7 M , confirming that the biosensor has excellent electroanalytical activity for phenol oxidation, with good stability and a wide linear range. Our electrode is based on a easily available and inexpensive material, as well as on its simple preparation, which has demonstrated high performance for phenol.
No takes yet. Share an insight, caveat, or question.
Moutcine et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: