This study demonstrates a novel conductive PCL bioelectrode for piezoresistive sensing in medical devices, highlighting its excellent durability and sensitivity.
Poly(ε‐caprolactone) (PCL) is a biodegradable polyester known for its low melting point and high flexibility, making it ideal for various medical device applications. In this study, we introduce a conductive nanocarbon black‐blended PCL as a reinforced ink for fabricating flexible and degradable piezoresistive bioelectrodes. This approach enables the creation of a piezoresistive bioelectrode optimized for precise biomechanical sensing. The bioelectrode exhibits exceptional mechanoelectrical stability under high tension (> 350%), repeated drawing (> 40%, 100 cycles), long‐term bending (> 7000 cycles), extrusion (> 10,000 cycles), and torsion (> 90°). When assembled into flexible piezoresistive sensors, the sensor achieves a high sensitivity (2.66 kPa −1 in the range of 0–3 kPa), along with excellent repeatability and durability (10,000 cycles at 5 N). The sensor has promising applications in human health monitoring, including finger, wrist, and elbow activity tracking, as well as knee joint space sensing for guiding precise surgical operations.
No takes yet. Share an insight, caveat, or question.
Wu et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: