Answers are grounded in this paper and cite their sources.
Ethylene oxide (Eto) is widely used as a sterilizing agent in industrial settings. It is a gas with a low vapour pressure and has the properties to penetrate deep in holes and cracks. It also kills almost all biological active material. It is therefore a perfect agent for sterilising heat and or gamma radiation sensitive medical equipment and other products. The method of sterilisation is described in a European Standard EN 550. The most common used method is sterilising by overkill. This is because Eto is cheap and with this method no costly bioburden tests are needed afterwards. The use of the overkill method has a negative aspect on the exposure of workers to Eto. Eto is a proven human carcinogen (category 1, IARC). Because Eto is a genotoxic carcinogen there is no safe exposure level. The Dutch exposure standard for Eto is currently 0,84 mg/m3. Resulting in an attributive cancer risk of 1 in 1000. Therefore the exposure to Eto should be reduced to the absolute minimum. Workers that enter the gas chambers after sterilisation and after a lag time use full face independent breathing protection. Therefore air monitoring is not suitable to measure there resulting exposure. To assess the remaining exposure, biological monitoring of a haemoglobin adduct (N-(2-hydroxy-ethyl)valine (HOEtVal) in red blood cells is used. It is a usable method with specific limitations. The method with its pitfalls and the results will be discussed as well as the positive effects on the remaining exposure.