In this study, an ultrasensitive label‐free electrochemical immunosensor based on Ti 3 C 2 T x (Mxene) with a 2D‐layered morphology has been proposed for the detection of EpCAM antigen. A hydrothermal method is employed for the synthesis of Ti 3 C 2 T x by using a less toxic exfoliating reagent, NaBF 4 . The bioelectrode, BSA/anti‐EpCAM/Ti 3 C 2 T x @ITO, has been fabricated by electrophoretic deposition of Ti 3 C 2 T x onto the ITO electrode, followed by an immobilization of EpCAM antibody. Electrochemical response studies reveal that the immunosensor shows a high sensitivity of 29.22 μA fg −1 ml cm −2 and a wide linear range from 0.1 fg/ml to 100 ng/ml, for EpCAM antigen detection. Further, the good stability of this fabricated immunosensor is an additional advantage for EpCAM antigen detection in serum samples.
No takes yet. Share an insight, caveat, or question.
Sweety et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: