Answers are grounded in this paper and cite their sources.
In present study, on a Jasco V-730 spectrophotometer, the method for estimating acetazolamide was used, with ethanol: distilled water (5:5) serving as the solvent. For acetazolamide, the recovery analysis research indicates recoveries of 102.55%, 102.68%, and 103.35% for various concentrations of the spiked medication. For quantifying acetazolamide in bulk and tablet dosage form, an RP-HPLC method was created and validated with mobile phase acetonitrile: water (80:20). For acetazolamide, the assay (% w/w) was found 98.6 %, 98.5 %, 98.9 %. The flow rate was 1ml/min. The sample volume was 20.00µl. Run time 5 min and retention time was 2.044.
No takes yet. Share an insight, caveat, or question.
Shruti Chougule (2024) studied this question.