Key result
Wearable cardiopulmonary sensor tracks clinical congestion markers in HF but fails to correlate with PCWP.
Why the study?
Does a novel non-invasive cardiopulmonary wearable sensor accurately measure congestion and correlate with established clinical measures of congestion in patients with heart failure or undergoing haemodialysis?
Cohort (n=66)
Does a novel non-invasive cardiopulmonary wearable sensor accurately measure congestion and correlate with established clinical measures of congestion in patients with heart failure or undergoing haemodialysis?
Answers are grounded in this paper and cite their sources.
A non-invasive wearable sensor worn for under 5 minutes successfully detected longitudinal fluid and congestion changes in hospitalized heart failure and haemodialysis patients, though single-point S3 measurements did not correlate with invasive PCWP.
Aims We examined the effectiveness of a novel cardiopulmonary management wearable sensor (worn for less than 5 mins) at measuring congestion and correlated the device findings with established clinical measures of congestion. Methods and results We enrolled three cohorts of patients: (1) patients with heart failure (HF) receiving intravenous diuretics in hospital; (2) patients established on haemodialysis, and (3) HF patients undergoing right heart catheterization (RHC). The primary outcomes in the respective cohorts were a Spearman correlation between (1) change in weight and change in thoracic impedance (TI) (from enrolment, 24 h after admission to discharge) in patients hospitalized for HF; (2) lung ultrasound B‐lines and volume removed during dialysis with device measured TI, and (3) pulmonary capillary wedge pressure (PCWP) and sub‐acoustic diastolic, third heart sound (S3) in the patients undergoing RHC. A total of 66 patients were enrolled. In HF patients ( n = 25), change in weight was correlated with both change in device TI (Spearman correlation [r sp ] = −0.64, p = 0.002) and change in device S3 (r sp = −0.53, p = 0.014). In the haemodialysis cohort ( n = 21), B‐lines and TI were strongly correlated before (r sp = −0.71, p < 0.001) and after (r sp = −0.77, p < 0.001) dialysis. Volume of fluid removed by dialysis was correlated with change in device TI (r sp = 0.49, p = 0.024). In the RHC cohort ( n = 20), PCWP measured at one time point and device S3 were not significantly correlated (r sp = 0.230, p = 0.204). There were no device‐related adverse events. Conclusions A non‐invasive device was able to detect changes in congestion in patients with HF receiving decongestion therapy and patients having fluid removed at haemodialysis. The cardiopulmonary management device, which measures multiple parameters, is a potentially useful tool to monitor patients with HF to prevent hospitalizations.
No takes yet. Share an insight, caveat, or question.
Curtain et al. (2024) conducted a cohort in Heart failure and haemodialysis congestion (n=66). Cardiopulmonary management wearable sensor vs. Established clinical measures of congestion (weight change, lung ultrasound B-lines, fluid volume removed during dialysis, and pulmonary capillary wedge pressure) was evaluated on Spearman correlation between (1) change in weight and change in thoracic impedance (TI) in hospitalized HF patients; (2) lung ultrasound B-lines and volume removed during dialysis with device TI; and (3) pulmonary capillary wedge pressure (PCWP) and device third heart sound (S3) in RHC patients (rsp = -0.64 (weight vs TI in HF); rsp = -0.71 pre-dialysis and -0.77 post-dialysis (B-lines vs TI); rsp = 0.49 (fluid volume vs TI); rsp = 0.230 (PCWP vs S3), p=0.002). The wearable cardiopulmonary sensor correlated with weight change in HF and B-lines pre/post-dialysis, but not with PCWP during RHC.
Synapse has enriched 10 related papers. Consider them for comparative context: