Background: Nowadays vital signs in clinics are electronically monitored excepted for urine output which still requires, manual recording, which is subject to human errors. We, in-vitro, assessed in this study the ability a novel electronic device to measure by a combination of two different methods, the real time flow-output, like urine output. Methods: Flow or urine output was measured by a combination technique of gravimetry using a dynamometer and capacitive sensing to improve precision. Data were obtained in 3 different situations, low, medium, and high flow. Results: Three different situations were tested: full or empty dead space and Pasteur’s chamber. The differences between the injected and measured volume varies from 0 to 5 ml representing a maximal error of 3 % regardless of the volume injected and the flow in ml/hour. The capacitive sensors behave as expected with a parallel curve of the expected and measured flow. Conclusion: We have shown in this in-vitro study that the combination of gravimetry and capacitive sensors analysis allow accurate measurement of volume and flow-output. Clinical trials should be conducted to confirm its usefulness in urine output measurement in the real life.