Answers are grounded in this paper and cite their sources.
Ultrasound image reconstruction from a single plane-wave transmission is required for many applications, However, imaging quality can be degraded when using conventional delay-and-sum (DAS) beamforming. This paper evaluates the performance of diffusion models (Diff) and conditional Generative Adversarial Networks (cGAN) for ultrasound image reconstruction when using the same base architecture, a UNet. Models were trained using a simulated dataset of 12500 acquisitions. Each sample featured a randomly positioned anechoic cyst in a medium with uniform sound speed, with downsampled IQ channel data serving as input. Results demonstrated that diffusion models could generate B-mode images of similar or improved contrast than the cGANs. On average, they exhibited a higher contrast-to-noise ratio (1.32 for Diff vs 1.11 for cGAN) and gCNR (0.83 for Diff vs 0.76 for cGAN).
No takes yet. Share an insight, caveat, or question.
Merino et al. (2024) studied this question.
Synapse has enriched 8 closely related papers on similar clinical questions. Consider them for comparative context: