Answers are grounded in this paper and cite their sources.
Arrhythmia detection is essential when assessing the safety of novel drugs and therapies in preclinical studies. Many short-term arrhythmia monitoring methods exist, including non-invasive ECG and Holter. However, there are no reliable, long-term, non-invasive, or minimally invasive methods for cardiac arrhythmia follow-up in large animals that allows free movement with littermates. A long follow-up time is needed when estimating the impact of long-lasting drugs or therapies, such as gene therapy. We evaluated the feasibility and performance of insertable cardiac monitors (ICMs) in pigs for minimally invasive, long-term monitoring of cardiac arrhythmias that allows free movement and species-specific behavior. Multiple implantation sites were tested to assess signal quality. ICMs recognized reliably many different arrhythmias but failed to detect single extrasystoles. They also over-diagnosed T-waves, resulting in oversensing. Muscle activity and natural startles of the animals caused noise, leading to a heterogeneous signal requiring post-recording evaluation. In spite of these shortcomings, the ICMs showed to be very useful for minimally invasive long-term monitoring of cardiac rhythm in pigs.
No takes yet. Share an insight, caveat, or question.
Airaksinen et al. (2024) studied this question.