Seven human donor eye globes underwent corneal cross‐linking using theranostic UV‐A device with accessory corneal iontophoresis system for patterned delivery of a 0.22% riboflavin solution. Theranostic‐guided UV‐A light illumination assessed riboflavin distribution and treated corneas at 10 mW/cm 2 for 9 min with a 5.0‐mm beam size. Corneal topography maps were taken at baseline and 2‐h post‐treatment. Analysis utilized corneal topography elevation data, with results showing controlled riboflavin delivery led to a consistent gradient, with 40% higher levels centrally (248 ± 79 μg/cm 3 ) than peripherally (180 ± 72 μg/cm 3 at ±2.5 mm from the center). Theranostic‐guided UV‐A light irradiation resulted in significant changes in corneal topography, with a decrease in best‐fit sphere value (−0.7 ± 0.2 D; p < 0.001) and consistent downward shift in corneal elevation map (−11.7 ± 3.7 μm). The coefficient of variation was 2.5%, indicating high procedure performance in achieving significant and reliable corneal flattening.
No takes yet. Share an insight, caveat, or question.
Lombardo et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: