Randomized trial shows active fluidics system affects visual outcomes in cataract surgery, suggesting optimal IOP levels.
To compare the central corneal thickness, mitigation events, post-operative vision, and complications in patients who underwent cataract surgery using the CENTURION ® Vision System with the ACTIVE SENTRY ® (Alcon Industries, Inc.) system and were assigned to one of two groups: intraocular pressure (IOP) group 20 or 50. Wavikar Eye Institute, Thane, India. A prospective comparative two arm, randomized, single blinded, interventional study. A total of 107 eyes were randomized to either the IOP-20 (n=55) or IOP-50 (n=52) group. All patients underwent phacoemulsification with an active fluidics system performed by a single surgeon; intraocular pressure (IOP) was set at either 20 or 50 mmHg. The mean [SD] baseline CCT in the IOP-20 group was 511.9 [35.7] μm and it was 521.1 [34.8] μm in the IOP 50 group (p=0.18). It had changed to 530.6 [39.8] μm in the IOP-20 group and 539.5 [39.2] μm in the IOP 50 (p=0.24). The mean [SD] ECD at the end of day 90 was 2423.0 [204.6] cells/mm2 in the IOP-20 group and 2476.1 [236.0] cells/mm2 in the IOP-50 group (p=0.22). Approximately 22% of the eyes in the IOP-20 group and 77% of the eyes in the IOP-50 group required active surge mitigation; this difference was statistically significant (p<0.001). We found that the phacoemulsification with an active fluidics system handpiece at 20 mmHg provided a safe and stable environment for cataract surgery.
No takes yet. Share an insight, caveat, or question.
Wavikar et al. (2025) studied this question.