This is the authors' abstract. We don't add key points for this paper.
Purpose: To assess the role of artificial intelligence (AI) based automated software for detection of Diabetic Retinopathy (DR) compared with the evaluation of digital retinography by two double masked retina specialists. Methods: Two-hundred one patients (mean age 65 ± 13 years) with type 1 diabetes mellitus or type 2 diabetes mellitus were included. All patients were undergoing a retinography and spectral domain optical coherence tomography (SD-OCT, DRI 3D OCT-2000, Topcon) of the macula. The retinal photographs were graded using two validated AI DR screening software (Eye Art TM and IDx-DR) designed to identify more than mild DR. Results: Retinal images of 201 patients were graded. DR (more than mild DR) was detected by the ophthalmologists in 38 (18.9%) patients and by the AI-algorithms in 36 patients (with 30 eyes diagnosed by both algorithms). Ungradable patients by the AI software were 13 (6.5%) and 16 (8%) for the Eye Art and IDx-DR, respectively. Both AI software strategies showed a high sensitivity and specificity for detecting any more than mild DR without showing any statistically significant difference between them. Conclusions: The comparison between the diagnosis provided by artificial intelligence based automated software and the reference clinical diagnosis showed that they can work at a level of sensitivity that is similar to that achieved by experts.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Musetti et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: