To minimize parallax errors and achieve high spatial resolution positron emission tomography (PET) systems, developing depth-of-interaction (DOI) encoding detectors has become a significant research topic. In this paper, we investigated a dual-ended readout PET detector based on the multi-voltage threshold sampling method combined with a convolutional neural network to calculate the pulse's energy (MVT-CNN method). The MVT sampling method was used to acquire time-threshold samples and digitize scintillation pulses. The CNN model was employed to establish an accurate mapping between MVT sampling points and energy information. The dual-ended readout detector's energy, DOI, and timing performance were evaluated with two irradiation configurations. The results demonstrated that the performance of the MVT-CNN method was close to that of the integration method based on oscilloscope sampling. Using the MVT-CNN method, the average energy resolution of the tested crystals over all depths was 14.5 ± 1.2%, and the average DOI resolution was 2.81 ± 0.70 mm. In the side irradiation configuration, the average coincidence timing resolution of the tested crystals at 2 mm depth was 435 ps. The performance of the dual-ended readout DOI-PET detector based on the MVT-CNN method suggested that it could develop small animal and organ-dedicated PET systems with high sensitivity and uniform spatial resolution.