We introduce a novel time-division multiplexing differential saturated absorption spectroscopy (TDMDSAS) approach, providing superior accuracy and stability in Doppler-free spectroscopy. By distinguishing probe and reference fields in the temporal domain, TDMDSAS efficiently suppresses Doppler broadening and common-mode optical noise. We utilized this technology to determine the absolute frequency of diverse neutral Yb isotopes across its 6s²\ ¹S₀→ 6s6p ¹P₁ transitions. Furthermore, the first-ever observation of in-situ Doppler-free Zeeman sub-level spectra was accomplished, enabling the determination of magnetic field gradients. We stabilized a UV diode laser at 399 nm using an error signal derived from the spectral first-derivative demodulated signal of ¹⁷⁴Yb. This technique yielded a frequency stability of up to 15 kHz with a 40 s averaging time and a standard deviation of around 180 kHz over a half-hour period. Given its low cost, straightforward, and scalable nature, TDMDSAS holds excellent potential in metrology and quantum applications.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: