This is the authors' abstract. We don't add key points for this paper.
Coherence scanning interferometry (CSI) is a widely used optical method for surface topography measurement of industrial and biomedical surfaces. The operation of CSI can be modeled using approximate physics-based approaches with minimal computational effort. A critical aspect of CSI modeling is defining the transfer function for the imaging properties of the instrument to predict the interference fringes from which topography information is extracted. Approximate methods, for example, elementary Fourier optics, universal Fourier optics, and foil models, use scalar diffraction theory and the imaging properties of the optical system to model CSI surface topography measurement. In this work, the simulated topographies of different surfaces, including various sinusoids, two posts, and a step height, calculated using the three example methods are compared. The presented results illustrate the agreement between the three example models.
No takes yet. Share an insight, caveat, or question.
Hooshmand et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: