Pilot study demonstrates high sensitivity of computer-aided detection for visible lesions in Barrett's esophagus, indicating potential utility during live endoscopic procedures.
Background and aims In this pilot study we evaluated performance of a recently developed computer-aided detection (CADe) system for Barrett's neoplasia during live endoscopic procedures. Methods 15 patients with and 15 without a visible lesion were included in this study. A CAD assisted workflow was employed that included: a slow pullback video recording of the entire Barrett's segment with live CADe assistance, followed by CADe assisted level-based video recordings every 2cm of the Barrett's segment. Outcomes were per patient and per level diagnostic accuracy of the CAD assisted workflow, where the primary outcome was per patient in-vivo CADe sensitivity. Results In the per patient analyses, the CADe system detected all visible lesions (sensitivity 100%). Per patient CADe specificity was 53%. Per-level sensitivity and specificity of the CADe assisted workflow were 100% and 73%, respectively. Conclusion In this pilot study, the CADe system detected all potentially neoplastic lesions in Barrett's esophagus comparable to an expert endoscopist. Continued refinement of the system may improve specificity. External validation in larger multicenter studies is planned