A sensitive and reliable LC–MS/MS method was developed and validated for the quantification of oxycodone and metabolites in human plasma. The method has a runtime of 6 min and a sensitivity of 0.1 μg/L for all analytes. Sample preparation consisted of protein precipitation. Separation was performed on a Kinetix biphenyl column (2.1 × 100 mm, 1.7 μm), using ammonium formate 5 m m in 0.1% aqueous formic acid and methanol LC–MS grade 100% in gradient elution at a flow rate of 0.4 ml/min. Detection was performed in multiple reaction monitoring mode using positive electrospray ionization. The method was linear over the calibration range of 0.1–25.0 μg/L for oxycodone, noroxycodone and noroxymorphone and 0.1–5.0 μg/L for oxymorphone. The method demonstrated good performance in terms of intra‐ and interday accuracy (86.5–110.3%) and precision (CV 1.7–9.3%). The criteria for the matrix effect were met (CV < 15%) except for noroxymorphone, for which an additional method was applied to compensate for the matrix effect. Whole blood samples were stable for 4 h at room temperature. Plasma samples were stable for 24 h at room temperature and 3 months at −20°C. Furthermore, the method was successfully applied in a pharmacokinetic drug interaction study of oxycodone and enzalutamide in patients with prostate cancer.