This abstract offers a method for monitoring post-operative temperature: an implantable antenna. This innovation in medical monitoring technology enables medical staff to continuously and remotely check on patients' core body temperatures without the use of probes or additional equipment. The proposed hexagonal microstrip patch antenna is printed on a biocompatible Alumina 95.99% substrate which has a thickness of 1.5mmThe planned antenna is intended to function in the range of 5.74 – 5.87 GHz in the ISM radio band with -34dB return loss. According to the SAR analysis, the highest SAR values were less than 1.6 W/Kg for 1g of tissue and 2 W/Kg for 10g of tissue , It is the Federal Communication Commission's (FCC) is advised maximum margin.