In this work, we use three different etching techniques—water bath, microwaves, and hotplate stirrer devices—in an effort to confirm the optimal etching time and conditions for tracks in the CR-39 solid-state nuclear track detector following irradiation to an alpha particle source (241Am). The research shows that the water bath’s maximum track density is 57596 tracks/cm2, appearing at 90 and 150 minutes. On the other hand, the hotplate stirrer’s maximum track density is 28050 tracks/cm2, appearing at 90 and 150 minutes. The microwave is the best etching method since it generated the tracks (65076 tracks/cm2) in a relatively short amount of time (10 min). The computation of the etching parameters, which are the bulk etch rates (V B ), track etch rate (V T ) and track etch ratio (V), was investigate in the etching properties. The time for the appearance of all traces of the microwave device is at a time of 10 minutes, followed by a hotplate stirrer at a time of 90 minutes, then a water bath at a time of 120 minutes.