In this work, we report the development of an optical sensor based on the colour variation of a silicon corrole upon interaction with specific emerging pollutants belonging to the PFAS family in water samples. The solid support on which the receptor is deposited consists of Colour Catcher® paper strips. An optical portable platform composed of low-cost electronic devices, such as an LED as a light source and a webcam as a detector, was developed to digitalize the strip colour changes during the measurements. This instrument is able to perform in situ analysis of water sources to determine the perfluoroalkyl substance (PFASs) content. Data analysis using the hue parameter allowed for the calculation, with great sensitivity, of the PFOA concentration depending on colour changes.