Background: Cervical disc arthroplasty (CDA) is currently used instead of fusion to preserve cervical spine motion. Cervical implant subsidence is a potential complication after CDA. Methods: Radiological measurements were recorded via patient anteroposterior and lateral radiographs in the neutral position. Subsidence was defined as a decrease of 3 mm or more in functional spinal unit height (FSUH) from which was measured on a post-operative (OP) radiograph. Results: This study included 104 patients who underwent 153 CDA levels with the Bryan Disc. Approximately one-quarter of the implants (22.9%) showed subsidence. Binary logistic regression analysis indicated that pre-OP mean disc height (DH) was identified as an independent risk factor for subsidence in multivariate analysis (0.151, 95% Confidence Interval 0–0.073, p = 0.018). Receiver operating characteristic curve analysis (area under the curve = 0.852, sensitivity 84.7%, specificity 77.1%) revealed a cut-off value of 4.48 mm for pre-OP Mean-DH in the risk for implant subsidence. Conclusions: In this study, the subsidence rate significantly increased when the implants were oversized beyond a pre-OP Mean-DH of approximately >4 mm. Moreover, the implant subsidence incidence was higher than that reported in previous studies. This is possibly due to endplate over-preparation or disc space over-distraction during placement at the same height as the Bryan Disc (8.5 mm).
No takes yet. Share an insight, caveat, or question.
Lee et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: