This is the authors' abstract. We don't add key points for this paper.
Chromatic Pupillometry, used to assess Pupil Light Reflex (PLR) to a coloured light stimulus, has regained interest since the discovery of melanopsin in the intrinsically photosensitive Retinal Ganglion Cells (ipRGCs). This technique has shown the potential to be used as a screening tool for neuro-ophthalmological diseases; however, most of the pupillometers available are expensive and not portable, making it harder for them to be used as a widespread screening tool. In this study, we developed a smartphone-based system for chromatic pupillometry that allows targeted stimulation of the ipRGCs. Using a smartphone, this system is portable and accessible and takes advantage of the location of the ipRGCs in the perifovea. The system incorporates a 3D-printed support for the smartphone and an illumination system. Preliminary tests were carried out on a single individual and then validated on eleven healthy individuals with two different LED intensities. The average Post-Illumination Pupil Light Response 6 s after the stimuli offsets (PIPR-6s) showed a difference between the blue and the red stimuli of 9.5% for both intensities, which aligns with the studies using full-field stimulators. The results validated this system for a targeted stimulation of the ipRGCs for chromatic pupillometry, with the potential to be a portable and accessible screening tool for neuro-ophthalmological diseases.
Answers are grounded in this paper and cite their sources.
No takes yet. Share an insight, caveat, or question.
Sousa et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: