This article proposes a slot-loaded circularly polarized implantable antenna with wide bandwidth at 2.45 GHz, i.e., Industrial, Scientific, and Medical (ISM) band. This antenna is established with a compact size of 44.52 mm 3 based on 3D printing technology, and biocompatible material MED610 is adopted for human safety consideration. During the design, unilaterally shorted patch antenna loaded with T-slot is adopted, obtaining two closely spaced orthogonal modes with miniaturized antenna size. More importantly, to obtain wide impedance matching, a spiral slot is introduced around the feed, forming an L -section network consisted of a parallel capacitor and a series inductor. As a result, a wide circular polarization bandwidth of 28.7 % is realized for the proposed antenna. The measured results are well agreed with simulated results.
No takes yet. Share an insight, caveat, or question.
Xu et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: