Due to GMR biosensor has the advantages of miniaturization, low-power consumption and rapid detection, so it has a widely application in bacterial, viral and cancer biomarker detection. However, the current GMR biosensor has relatively low detection sensitivity in biotarget detection, that limiting the development in biomedical detection applications rapidly. As we know, the magnetic flux concentrator (MFC) can improve detection sensitivity and resolution of the GMR sensor. So, a new micro-funnel-shaped MFC is proposed for GMR biosensor and analyzed by finite element method (FEM). The MFC shape, size and distribution of magnetic labels were investigated. The simulation results indicate that the signal gain can be increased by 30% with the hypotenuse of MFC. The distribution of magnetic labels was arranged horizontally along the X-axis can generate high magnetic field intensity and the strongest magnetic gain. The simulation results has certain theoretical guidance for high sensitivity GMR biosensors design.
No takes yet. Share an insight, caveat, or question.
Sun et al. (2024) studied this question.