Background Prostatic artery embolization (PAE) is a safe, minimally invasive angiographic procedure that effectively treats benign prostatic hyperplasia; however, PAE-related patient radiation exposure and associated risks are not completely understood. Purpose To quantify radiation dose and assess radiation-related adverse events in patients who underwent PAE at multiple centers. Materials and Methods This retrospective study included patients undergoing PAE for any indication performed by experienced operators at 10 high-volume international centers from January 2014 to May 2021. Patient characteristics, procedural and radiation dose data, and radiation-related adverse events were collected. Procedural radiation effective doses were calculated by multiplying kerma-area product values by an established conversion factor for abdominopelvic fluoroscopy-guided procedures. Relationships between cumulative air kerma (CAK) or effective dose and patient body mass index (BMI), fluoroscopy time, or radiation field area were assessed with linear regression. Differences in radiation dose stemming from radiopaque prostheses or fluoroscopy unit type were assessed using two-sample