Purpose: To characterize corneal biomechanical properties using the CORVIS-ST device in myopic individuals.Methods: This prospective cross-sectional study included patients with myopia.Our study included 154 eyes of 154 myopic patients aged between 18 and 40 years, with stable refraction for at least 2 years.A full ophthalmological examination and corneal tomography were performed using a Pentacam HR device.Corneal biomechanical parameters were assessed using the CORVIS-ST device in mild, moderate, severe, and extreme myopia groups.Results: Statistically significant differences were observed in the DA ratio (p = 0.033), SP-A (p=0.009),CBI (p=0.041),SSI (p=0.000), and Peak distance (p = 0.032).In correlation with different Corvis ST biomechanical variables, SE was found to be correlated with DA ratio(r=-0.191,p=0.018),SP-A(r=0.199,p=0.013) and SSI(r=-0.336,p=0.000), while in multiple regression analysis, SE was found to be independently correlated with SSI and peak distance(p=0.036,0.038respectively) while the grade of myopia was found to be independently correlated with SP-A(p=0.034).Conclusion: SSI, Peak distance, and SP-A were independently related to SE and myopia grade, confirming the hypothesis that eyes with higher myopia are more deformable and less stress resistant.