This study investigates the charging-up effect on the Topmetal-II⁻ chip in Gas Micro-Pixel Detectors(GMPD). It is found that this effect differs from the charging-up typically observed in gas detector multiplier devices and increases the effective gain of the detector. The research indicates that this effect originates from the accumulation of charges on the insulating layer of the chip's pixel surface and is unlikely to dissipate naturally. Iterative simulations using COMSOL and GARFIELD++ were employed to model the variation of detector effective gain with the charging-up effect, and a simple yet effective model was proposed, which aligns well with experimental data. The study suggests that to ensure stable detector gain and reduce residual modulation in polarization detection, charging-up to near saturation state is required before operating the GMPD.
No takes yet. Share an insight, caveat, or question.
Yi et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: