Purpose: Basilar thumb arthritis is a common condition that can be associated with significant disability. While surgical management has historically consisted of trapeziectomy-type procedures, hemiarthroplasty has become an increasingly common alternative. While outcomes have been promising, they vary according to the implant used. We sought to evaluate the performance of the BioPro® Modular Thumb implant. Methods: This was a retrospective review of 110 thumbs with carpometacarpal arthritis that underwent hemiarthroplasty between the years 2008 and 2016 with the BioPro® device. All thumbs had Eaton-Little Stage II or III arthritis. Patients were asked to complete preoperative and postoperative assessments using QuickDASH, as well as complete strength measurements using a dynamometer. Results: Mean age was 61 years, with average final follow-up at 3.8 years. Forty thumbs completed QuickDASH and showed a 34.78-point improvement at average 2-year follow-up. Implant survivorship was 88.18% at average 3.8-year follow-up. The most common cause of revision was titanium metal allergy. No implants dislocated. Conclusion: Outcomes of hemiarthroplasty using the BioPro® device were comparable or superior to alternatives and support its continued use.
No takes yet. Share an insight, caveat, or question.
Goodsett et al. (2024) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: