Answers are grounded in this paper and cite their sources.
Introduction Current wearables that collect heart rate and acceleration were not designed for children and/or do not allow access to raw signals, making them fundamentally unverifiable. This study describes the creation and calibration of an open-source multichannel platform (PATCH) designed to measure heart rate and acceleration in children ages 3-8 years. Methods Children (N = 63; Mean age = 6.3) participated in a 45-minute protocol ranging in intensities from sedentary to vigorous activity. Actiheart-5 was used as a comparison measure. We calculated mean bias, mean absolute error (MAE) mean absolute percent error (MA%E), Pearson correlations and Lin’s concordance correlation coefficient (CCC). Results Mean bias between PATCH and Actiheart heart rate was 2.26 BPM, MAE was 6.67 BPM and M%E was 5.99%. The correlation between PATCH and Actiheart HR was .89 and CCC was .88. For acceleration, mean bias was 1.16 mg and MAE was 12.24 mg. The correlation between PATCH and Actiheart was .96 and CCC was .95. Conclusions The PATCH demonstrated clinically acceptable accuracies to measure heart rate and acceleration compared to a research grade device.