Answers are grounded in this paper and cite their sources.
Lactate serves as a crucial biomarker that indicates sepsis assessment in critically ill patients. A rapid, accurate, and portable analytical device for lactate detection is required. This work developed a stepwise polyurethane–polyaniline–m-phenylenediamine via a layer-by-layer based electrochemical biosensor, using a screen-printed gold electrode for lactate determination in blood samples. The developed lactate biosensor was electrochemically fabricated with layers of m-phenylenediamine, polyaniline, a crosslinking of a small amount of lactate oxidase via glutaraldehyde, and polyurethane as an outer membrane. The lactate determination using amperometry revealed the biosensor’s performance with a wide linear range of 0.20–5.0 mmol L−1, a sensitivity of 12.17 ± 0.02 µA·mmol−1·L·cm−2, and a detection limit of 7.9 µmol L−1. The developed biosensor exhibited a fast response time of 5 s, high selectivity, excellent long-term storage stability over 10 weeks, and good reproducibility with 3.74% RSD. Additionally, the determination of lactate in human blood plasma using the developed lactate biosensor was examined. The results were in agreement with the enzymatic colorimetric gold standard method (p > 0.05). Our developed biosensor provides efficiency, reliability, and is a great potential tool for advancing lactate point-of-care testing applications in the early diagnosis of sepsis.
No takes yet. Share an insight, caveat, or question.
Thongkhao et al. (2024) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: