Answers are grounded in this paper and cite their sources.
A novel method determines p-nitroaniline and its metabolites in blood, highlighting excellent accuracy and reliability.
An accurate, reliable, and sensitive method was successfully developed to simultaneously determine p-Nitroaniline (p-NA) and its metabolites in blood based on the hyphenated technology of quadrupole-orbitrap high resolution mass spectrometry (Q-Orbitrap HRMS) and ultra-performance liquid chromatography (UPLC). After the blood sample was extracted with ethyl acetate, the organic layer was taken for injection and analysis. Within the concentration range of 1–100 μg/L, the calibration curves of p-NA and its metabolites had a good linearity, with correlation coefficient( r ) values greater than 0.999. This method has excellent precision and accuracy. The intra- and inter-day coefficients of variation (CVs) were less than 9.9% and 8.7% respectively, and the analytical accuracy ranged from 83.1% to 101.3%. The lower limits of detection (LLODs) for all target analytes were between 0.6 μg/L and 2.2 μg/L, and the lower limits of quantification (LLOQs) were between 2.0 μg/L and 7.4 μg/L. In addition, the application potential of this method was verified by analyzing the blood samples of workers exposed to p-NA. The results indicated that this method was accurate, reliable and sensitive, and was applicable to the detection of p-NA and its metabolites in the blood.
No takes yet. Share an insight, caveat, or question.
Wang et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: