This technique details how robotic assistance improves bone preservation and ligament balance in TKA conversion from UKA. It highlights the importance of accurate planning and execution.
Converting a medial unicompartmental knee arthroplasty (UKA) to total knee arthroplasty (TKA) presents challenges in bone preservation, alignment, and soft-tissue balance. Robotic assistance enables three-dimensional CT-based planning, precise bone preparation, and real-time ligament balance assessment, thereby supporting a bone-preserving approach. We describe a stepwise workflow using the Mako system (Stryker) to convert a failed medial UKA to a condylar-stabilizing (CS) Triathlon TKA within a functional alignment framework. Pre-explantation registration on the in-situ components maintains accuracy despite potential metallic artifacts. The polyethylene insert is briefly removed for reference surface acquisition, then reinserted for intraoperative balance evaluation. After component removal, small medial tibial or femoral defect can be filled with autologous cancellous bone graft from the resected bone surfaces. Definitive cementless components are implanted without stems or augments, and the patella is resurfaced. This technique provides a reproducible robotic workflow for UKA-to-TKA conversion in selected cases with preserved bone stock and stable soft-tissue balance, and facilitates accurate and reproducible conversion with optimal bone preservation.
No takes yet. Share an insight, caveat, or question.
Mahlouly et al. (2025) studied this question.