October 10, 2025Advanced Science5 citationsOpen Access

Long‐Term Stable Subdural Recordings Enabled by Fibrosis‐Resistant Hydrogel‐Integrated µECoG Arrays

View Full Paper
LCLin ChenHZHao ZhongLWLei Wang

Key Points

  • aGel-µECoG maintained 94.8% signal fidelity over 16 weeks, vastly outperforming conventional arrays.
  • Histological analyses confirmed a 20-fold reduction in impedance increases due to less neuroinflammation and fibrosis.
  • The hydrogel facilitates sutureless implantation, enhancing mechanical compatibility with brain tissue.
  • The adhesive hydrogel-µECoG array allows for long-term electrical communication across chronic applications.

Abstract

Abstract The long‐term performance of microscale subdural arrays is often compromised by adverse tissue responses, leading to increased electrochemical impedance and signal deterioration during chronic applications. Here, the study presents an adhesive hydrogel‐integrated micro‐electrocorticography (aGel‐µECoG) array that improves mechanical compatibility and adhesion to brain tissue, effectively mitigating tissue responses and enabling stable, high‐performance electrical communication over days to months. The hydrogel consists of a hydrophilic polyvinyl alcohol and hydrophobic poly3‐(trimethoxysilyl) propyl methacrylate heteronetwork, serving as a biological and mechanical bridging to enable seamless, anti‐fibrotic, long‐term integration with brain tissue. The aGel‐µECoG achieves robust tissue adhesion (25.2 ± 3.8 kPa) with reversible and safe removal, enabling sutureless implantation. With an ultrathin 10‐µm ionic conductive hydrogel coating, the array exhibits high electrical fidelity, effectively preserving fine‐grained sub‐millimeter spatial resolution and maintaining unattenuated signal strength for functional cortical mapping. Compared to conventional µECoG arrays, aGel‐µECoG exhibits a 20‐fold reduction in acute‐phase impedance increases, attributed to significantly reduced neuroinflammation and fibrotic tissue formation, as confirmed by histological analyses. Long‐term recordings further reveal that aGel‐µECoG maintained 94.8% of the signal‐to‐noise ratio for steady‐state visually evoked potentials over 16 weeks, whereas uncoated µECoG arrays declined to 69.5%. These findings establish aGel‐µECoG as a durable and high‐performance neural interface.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Chen et al. (2025) studied this question.

synapsesocial.com/papers/68e8619c7ef2f04ca37e421fhttps://doi.org/10.1002/advs.202515453