Synapse
⌘+K
Synapse
PulseExploreClubsResearchersJournals
Instagram
HomeClubsExplore
October 18, 2025OtolaryngologyOpen Access

Early Failure Risk and Long‐Term Outcomes of Advanced Bionics Electrode Devices Post‐Recall

View Full Paper
Ask AI
Bookmark
Share

Authors

KLKalena LiuLOLauren O’BrienDCDavid Campbell

Discussion

Loading...

Member takes

Overview

Retrospective cohort study assessed electrode failure in Advanced Bionics implants, suggesting long-term stability.

Key Points

  • Device failures in Advanced Bionics cochlear implants mostly happen within the first 5 years post-implantation.
  • In total, 24 out of 140 implants (17.1%) experienced failure, with a median failure time of 3.6 years.
  • Kaplan-Meier survival analysis showed a significant decrease in failure rate after the first 5 years, down to 2.1% annually.
  • Findings highlight long-term stability of cochlear implants, aiding in patient counseling and follow-up care.

Cite This Study

Liu et al. (2025) studied this question.

synapsesocial.com/papers/68f35bfc73f0a7d050f47e0ahttps://doi.org/10.1002/ohn.70048