Novel transceiver array improves sodium and proton imaging in brain MRI, suggesting enhanced workflow efficiency.
With the increasing adoption of ultra‐high‐field (UHF) systems, sodium ( 23 Na) imaging is undergoing clinical translation. However, existing commercially available dual‐tuned coils, while demonstrating adequate sodium signal‐to‐noise ratio (SNR), fail to meet the requirements for high‐resolution proton ( 1 H) structural imaging, resulting in suboptimal workflow efficiency during integrated multi‐nucleus examinations. In this work, we present a novel double‐layer 1 H/ 23 Na transceiver array with eight channels per nucleus for human brain MRI at 7T. The proposed array achieved 64% of the 1 H SNR of the 32‐channel receive array commonly used in clinical practice. By integrating dual‐nucleus imaging capabilities within a single platform, this design eliminates coil‐switching requirements and post‐acquisition registration, thereby streamlining clinical workflows and advancing sodium MRI translational feasibility.
No takes yet. Share an insight, caveat, or question.
Lou et al. (2025) studied this question.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: