October 19, 2025Brazilian Journal of Radiation SciencesOpen Access

Monte Carlo simulation of a 6 MV Varian Trilogy clinical linear accelerator using Geant4 for small-field dosimetry analysis

View Full Paper
Ask AI
Bookmark
Share

Authors

ARA. L. B. RibeiroAOAlex Cristóvão Holanda de OliveiraLSLeonardo P. Silva

Discussion

Loading...

Member takes

Overview

Monte Carlo simulation demonstrates accurate dose profiles in 6 MV linear accelerators, highlighting challenges in small-field dosimetry.

Key Points

  • Simulation results showed a 2% difference in depth dose curves compared to experimental measurements.
  • Over 95% of points passed the gamma analysis for a 10 x 10 cm² field, confirming the simulation's reliability.
  • A Geant4-based framework facilitates correction factors for small-field dosimetry, addressing electronic disequilibrium.
  • Preliminary data reveals discrepancies in output factors for small photon fields, indicating a need for specific corrections.

Cite This Study

Ribeiro et al. (2025) studied this question.

synapsesocial.com/papers/68f43eeb854d1061a58ab839https://doi.org/10.15392/2319-0612.2025.2958