Diagnostic Performance of CBCT in Detecting Different Types of Root Fractures with Various Intracanal Post Systems

View Full Paper
Ask AI
Bookmark
Share

Authors

SESerhat EfeoğluESEcem SiskoAÖAyşenur Öncü

Discussion

Loading...

Member takes

Overview

Analysis shows diagnostic accuracy in detecting root fractures in teeth, suggesting CBCT device choice affects outcomes.

Key Points

  • NewTom 7G demonstrated superior diagnostic accuracy for root fractures, particularly at smaller voxel sizes.
  • Observers showed almost perfect agreement with the gold standard in fracture detection, achieving κ ≥ 0.81.
  • Controlled fractures were induced using a universal testing machine, assessing multiple imaging modalities across two CBCT devices.
  • Selecting the right CBCT device and imaging parameters is crucial for accurate detection of vertical root fractures.

Cite This Study

Efeoğlu et al. (2025) studied this question.

synapsesocial.com/papers/68f9f86eb2c35e10cc4e3be1https://doi.org/10.3390/tomography11100116
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1In Vitro Detection of Dental Root Fractures with Cone Beam Computed Tomography (CBCT)2014 · 13 citations
  2. 2Comparison of imaging methods for detection of simulated horizontal and oblique root fractures in a new study model2025
  3. 3Dental undergraduate students' ability in diagnosing horizontal root fractures: a comparative analysis using CBCT and periapical radiography2026
  4. 4Impact of <scp>CBCT</scp> acquisition protocols and artifact reduction filters on diagnosing vertical root fractures2024 · 4 citations
  5. 5Diagnostic Accuracy of <scp>CBCT</scp> Versus <scp>2D</scp> Radiography for Various Dental Pathologies: A Systematic Review and Meta‐Analysis2026