October 27, 2025European Journal of Cardio-Thoracic Surgery4 citationsOpen Access

Low driveline infection rates in patients with a novel fully magnetically levitated ventricular assist device

View Full Paper
XWXian-Qiang WangXZXingtong ZhouHCHaibo Chen

Key Points

  • Driveline infections occurred in 20 of 181 patients, representing an 11.0% rate during a median support duration.
  • The cumulative incidence of first driveline infection reached 8.5% at 24 months, indicating effective management strategies.
  • Data were retrospectively analyzed from patients who underwent CH-VAD implantation across multiple centers from June 2017 to October 2024.
  • Increased dressing changes were associated with a significant rise in infection risk, highlighting the need for careful management.

Abstract

Abstract OBJECTIVES Driveline infection remains a major complication of long-term ventricular assist device therapy. Previous studies have associated the risk of infection with driveline design and physical properties. The CH-VAD is a fully magnetically levitated pump with an ultrathin (diameter 3.3 mm), highly flexible driveline designed to reduce infection risk. This study was conducted to assess the incidence, risk factors, approaches to prevention and management of driveline infection in CH-VAD patients (NCT06878456). METHODS All patients who underwent CH-VAD implantation between June 2017 and October 2024 at nine centers were retrospectively analyzed. Data related to clinical characteristics, surgical procedures, driveline infection prevention and management details were collected, and risk factors were analyzed. RESULTS A total of 181 consecutive patients were included, with a median age of 56 years and 86% male. During a median support duration of 1.6 years, driveline infections occurred in 20 out of 181 patients (11.0%). Twenty-six events were observed, corresponding to a rate of 0.07 events per patient-year. Cumulative incidence of the first driveline infection was 2.8% at 6 months, 5.6% at 12 months and 8.5% at 24 months. Frequent dressing changes were found to be significantly associated with increased driveline infection risk (HR 3.773, p = 0.022). CONCLUSIONS Patients treated with CH-VADs experienced an encouraging low rate of driveline infection, considered among the lowest reported for durable ventricular assist devices. These findings have implications for both clinical management and driveline design.

Ask AI
Helpful
Bookmark
Share
View Full Paper

Cite This Study

Wang et al. (2025) studied this question.

synapsesocial.com/papers/68ff87d8c8c50a61f2bdcbb8https://doi.org/10.1093/ejcts/ezaf365
Ask AI
Helpful
Bookmark
Share
View Full Paper

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Left ventricular assist systems and infection-related outcomes: A comprehensive analysis of the MOMENTUM 3 trial2020 · 101 citations
  2. 2Mechanical Characteristics and Infection Risks: Experimental Insights and Clinical Outcome Prediction on Left Ventricular Assist Device Driveline Features2025 · 1 citations
  3. 3The Importance of the Management of Infectious Complications for Patients with Left Ventricular Assist Device2015 · 16 citations
  4. 4Reduction in driveline infection rates: Results from the HeartMate II Multicenter Driveline Silicone Skin Interface (SSI) Registry2014 · 82 citations
  5. 5Updated definitions of adverse events for trials and registries of mechanical circulatory support: A consensus statement of the mechanical circulatory support academic research consortium2020 · 228 citations