This study demonstrates improved diagnostic performance for retinal tortuosity in premature infants, indicating software can enhance ROP assessments.
This study evaluated the diagnostic performance of a novel vessel tracing software that quantifies retinal vessel tortuosity in premature infants for assessing retinopathy of prematurity (ROP). Seventy-three retinal images of premature infants were subjectively classified as having "Normal" or "Abnormal" tortuosity based on arterial tortuosity of the posterior retina by seven clinicians. The software then semi-automatically traced the same arteries within circular regions of interest centered at the optic disc, with radii ranging from 2 to 20x the disc diameters. Vessel tortuosity was defined as the sum of external angles at vessel bend points. Tortuosity was significantly greater in Abnormal images within regions exceeding 4-disc diameters (p < 0.05). Diagnostic accuracy improved with larger regions: the area under the curve increased from 0.56 at 2 times to 0.90 at 8 times and 1.00 at 14 times the diameter. At 14 times, corresponding approximately to Zone I, the software achieved 100% sensitivity and specificity for detecting abnormal tortuosity. The software accurately quantified vessel tortuosity especially within Zone I of the retina, which may reduce inter-observer variability and aid clinicians, including non-specialists, in the evaluation of ROP eyes.
No takes yet. Share an insight, caveat, or question.
Zhou et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: