Answers are grounded in this paper and cite their sources.
This analysis reveals patient errors as a superior risk metric compared to sigma, indicating its relevance for risk management.
Introduction/Objective Sigma metrics, calculated as [(TEa - |bias|)/SD], are frequently used to estimate errors per million results in analytical processes. Problems relative to risk management include: 1. Sigma measures proximity to the nearest allowable error limit but underestimates errors when bias is minimal or absent. 2. Sigma does not vary with the number of patient samples analyzed annually. Methods/Case Report We calculated sigma for 224 routine chemistry QC samples from four instruments across three laboratories based on January 2025 data. We also used software from ElevateQC (Toronto, Ontario, Canada) to calculate sigma metrics and the number of results outside both upper and lower TEa limits annually based on patient volumes. Results QC samples represented between 3,650 and 30,660 patient results annually. In one case, a 14% increase in sigma (0.83 to 0.95) led to a 724% increase in annual errors (729 to 5,279.) Similarly, a 23% increase in sigma (0.83 to 1.02) resulted in a 425% increase in errors (729 to 3,099). The correlation coefficient between sigma and errors per year was only 0.18. Conclusion Errors per year is a more informative metric than sigma, especially when communicating patient risk levels to non-scientists.
No takes yet. Share an insight, caveat, or question.
Brooks et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: