Answers are grounded in this paper and cite their sources.
Case report reveals titanium miniplate fracture associated with bruxism in a 54-year-old woman after internal fixation.
Fracture of titanium miniplates after internal fixation of mandibular condylar fractures is rare, accounting for 2.6% of cases in our previous study. A 54-year-old woman presented with a right condylar neck fracture with medial displacement from a fall due to alcohol. Internal fixation was performed with two 4-hole titanium miniplates with nonlocking screws. Postoperative images showed anatomic reduction without deviation. No intermaxillomandibular fixation was performed, and the patient followed a soft diet for 3 weeks. A panoramic radiograph at 5 months postoperatively showed a miniplate fracture. The patient’s maximum mouth opening was 45 mm, with no malocclusion or symptoms. Scanning electron microscopy revealed striation lines on the removed, fractured hole, indicating a metal fatigue fracture. Postoperative interviews revealed the patient had bruxism and failed to follow the no smoking protocol.
No takes yet. Share an insight, caveat, or question.
Sasaki et al. (2025) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: