Analysis shows carbon nanotubes enhance electrical conductivity in flexible EEG electrodes, indicating improved brain activity monitoring.
This study evaluates styrene–ethylene–butylene–styrene (SEBS) composites modified with carbon nanotubes (CNTs) and carbon black (CB) for flexible electroencephalography (EEG) electrodes. Maleic anhydride‐grafted SEBS (SEBS‐MA) with 8 wt% CNTs and 2 wt% CB provides an optimal balance of conductivity and flexibility, a storage modulus comparable to SEBS, and yields a highly flexible conductive material. SEBS and SEBS‐MA composites with 8 wt% CNT/2 wt% CB produced stable, low‐noise signals, suggesting responsiveness to brain activity. The contact impedance of the elastomeric thermoplastic polymer (SEBS)/8 wt% CNT/2 wt% CB electrode is 4.25 ± 0.5 kΩ, and 4.5 ± 0.6 kΩ for SEBS‐MA/8 wt% CNT/2 wt% CB, comparable to a commercial electrode (4.75 ± 1.5 kΩ). SEBS/8 wt% CNT/2 wt% CB and SEBS‐MA/8 wt% CNT/2 wt% CB produced stable, low‐noise EEG signals. In vivo EEG recordings demonstrated that SEBS‐MA with 8 wt% CNT/2 wt% CB effectively captured transitions between the eyes‐open and eyes‐closed states, yielding clear and stable signals. These findings suggest that SEBS‐MA/8 wt% CNT/2 wt% CB is a promising material for flexible, high‐performance EEG electrodes due to its balance of electrical conductivity, mechanical stability, and signal clarity.
No takes yet. Share an insight, caveat, or question.
Ilyass Tabiai (2025) studied this question.